Provide an INF file that specifies the Ports device setup class and the Serial function driver as the service for the port. To configure a serial port as COM port, comply with the requirements that are defined in Configuration of COM Ports. American medical port devices driver.
Documentation
Operator's Manual - WristOx2 3150 with USB View PDF
Utilities and Drivers - WristOx2 3150 with USB View Folder
Operating Instructions - 3150 Memory Clear Application View PDF
WristOx2 3150 Pulse Oximeter: Use/Battery Change View Video
Spec Sheet - WristOx2 Model 3150 USB View PDF
Disposable Wrist Strap Instructions - WristOx2 Model 3150 USB View PDF
WristOx2 for Sleep Studies View PDF
WristOx2 for Ambulatory Monitoring View PDF
Sensor Compatibility Guide View PDF
Resources
Understanding Accuracy, Bias, and Variability in SpO2 Readings View
PureSAT SpO2 Averaging View
Nonin Medical’s WristOx2® Model 3150 Beats VirtuOx VPOD in Hypoxia Testing View
Accuracy and Superior Performance of PureSAT and PureLight Oximetry Technologies View
Using Overnight Pulse Oximetry to Manage Oral Appliance Therapy (OAT) during Treatment for Obstructive Sleep Apnea View
Accurate Oximetry Monitoring – See the Technology Difference View
Pulse Oximetry at Home View
WristOx2® Model 3150 with USB Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Displays:
Numeric: 3-digit LCD
Pulse Strength: Pulse Strength Bar Graph
Accuracy – Sensors: Declared accuracy data for compatible sensors can be found in Nonin’s Sensor Accuracy document.
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW maximum average
Infrared: 910 nanometers @ 1.2 mW maximum average
** This information is especially useful for clinicians performing photodynamic therapy.
Temperature:
Operating: -5 °C to 40 °C (23 °F to 104 °F)
Storage/Transportation: -40 °C to 70 °C (40 °F to 158 °F)
Time (from storage) for monitor to be ready for its intended use: 10 minutes to warm from -40 °C to -5 °C
10 minutes to cool from 70 °C to 40 °C
Device temperature will not exceed 41°C as measured during a controlled environment test.
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Operating Altitude: Up to 4,000 meters (13,123 feet)
Operating Hyperbaric Pressure: Up to 4 atmospheres
Power Requirements: Two AAA (1.5V) batteries
Battery Life (expected minimum): NOTE: Based on testing new and fully-charged batteries. See footnotes for brands used. Refer to battery manufacturers’ operator’s manuals for instructions for use. | Alkaline AAAa) | Rechargeable AAA (700 mAh)b | Rechargeable AAA (1050 mAh)c |
---|---|---|---|
Storage: MVI Display Mode off: MVI Display Mode on: | 9 months 25 days | Not specified | Not specified |
Operating without Bluetooth Connection, continuous use: | 53 hours | 36 hours | 52 hours |
Operating with Bluetooth Connection, continuous use: | 44 hours | 24 hours | 31 hours |
Dimensions (without sensor or wristband): 51 mm x 73 mm x 19 mm (H x W x D)
(2.0 in. x 2.9 in. x 0.75 in.)
Weight (with batteries and wristband): 70.0 g (2.5 oz)
Memory:
Type: Non-volatile
Capacity:
up to 1,080 hours (4 sec. data storage rate)
up to 540 hours (2 sec. data storage rate)
up to 270 hours (1 sec. data storage rate)
Classification per ANSI/AAMI ES60601-1 and CAN/CSA-C22.2 No. 60601-1:
Type of Protection: Internally powered (battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection: IP33
This product complies with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation
and testing.
a.Batteries used: Harding Alkaline AAA
b.Batteries used: Energizer (MR03) Ni-MH 700mAh Rechargable Batteries 1.2 VDC
c.Batteries used: Ansmann (HR03) Ni-MH 1050 nMh Rechargable Batteries 1.2 VDC
WristOx2® Model 3150 with USB Troubleshooting
Problem | Possible Cause | Possible Solution | |||
Device will not activate. | Batteries inserted wrong. | Check batteries. | |||
Batteries are depleted. | Replace batteries. | ||||
Sensor is disconnected. | Reconnect sensor. | ||||
Device is in Sensor Activation mode and has timed out. | Press the activation switch. | ||||
Disconnect and then reconnect the sensor. | |||||
Device is in Programmed mode. | Use nVISION software to select Spot Check or Sensor Activation mode. | ||||
%SpO2 and pulse rate do not display. | Device set in Partial Display mode. | Use nVISION software to select Full Display mode. Reconnect sensor. | |||
Poor pulse signal indicator displays. | Excessive patient motion. | Reduce patient motion. | |||
Poor pulse signal indicator displays and pulse strength indicator shows two bars or less. | Inadequate pulse signal. | Reposition or replace sensor, or place sensor on a different finger. | |||
Remove and reconnect sensor. | |||||
Hands are cold. | Warm sensor application site. | ||||
No pulse display on pulse strength bar graph indicator. | Sensor applied incorrectly. | Refer to sensor Instructions for Use for proper sensor application. | |||
Device needs repair. | Contact Nonin Technical Service. | ||||
Possible interference from blood flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.). | Reduce or eliminate restriction. | ||||
Reduced circulation due to excess pressure from sensor. | Check sensor alignment, reposition sensor, verify correct sensor size. | ||||
Excessive ambient light. | Shield sensor from light source. Check sensor alignment. | ||||
Problem | Possible Cause | Possible Solution | |||
No pulse display on pulse strength indicator (continued). | Sensor applied to polished or artificial nail. | Remove fingernail polish or an artificial nail. | |||
Sensor Light-Emitting Diode (LED) is not lit. | Contact Nonin Technical Service. | ||||
Er 01 displays on LCD. | Device configuration memory failure. | Device reverts to default settings (Spot-Check mode, 4-second sample rate). Use nVISION software to change settings. If error code continues, contact Nonin Technical Service. | |||
Er 02 or 04 displays on LCD. | Device memory failure. | Contact Nonin Technical Service. | |||
Er 03 or 05 displays on LCD. | Device failure. Device memory intact, but device may have lost most recent session or stored data. | If error code continues, contact Nonin Technical Service. | |||
Dashes continually display on LCD. | Sensor malfunction. | Replace sensor with a Nonin-branded sensor. | |||
Device does not record in Programmed mode. | Data collection start and stop times are set incorrectly. | Use nVISION software to program correct start and stop times. | |||
Clock settings are lost after replacing batteries. | Use nVISION software to reset clock. | ||||
Devices will not pair. | Device is out of range. | Verify device is in range while being paired (approximately 100 meters [328 feet] spherical radius). | |||
Bluetooth radio has timed out. | Press activation switch to turn on Bluetooth radio. |
If these solutions do not correct the problem, please contact Nonin Technical Service at (800) 356-8874 (USA and Canada), + 1 (763) 553-9968, or +46 650 401500 (Europe).
Customer Support
For general questions, please contact Nonin Customer Support.
Technical Support
For product technical, troubleshooting or repair questions, please contact Nonin Technical Support.
Subscribe
Stay informed with updates from Nonin
Nonin Medical Inc
(Global Headquarters)
Address
13700 1st Ave N,
Plymouth, MN 55441
Email
info@nonin.com
Toll Free (USA and Canada)
1.800.356.8874
Phone
1.763.553.9968
Nonin Medical B.V.
(Europe)
Address
Doctor Paul Janssenweg 150
5026 RH Tilburg
The Netherlands
Email
infointl@nonin.com
Technology Leadership
Support to help our OEM partners succeed
Nonin Medical is a leading global provider of OEM solutions in pulse and regional oximetry. Our easy-to-integrate products and dedicated customer support team have helped more than 200 OEM customers bring more than a million low power, high accuracy medical devices to market. Nonin has one of the world’s largest installed bases of oximetry components in medical device solutions used by professionals in homecare, sleep medicine, hospitals, EMS, and telehealth.
A History
of Innovation
Nonin Medical Port Devices Driver Download Windows 7
We created the world’s first low power pulse oximeter for OEM applications. Our proven technologies provide consistent performance in a broad range of our partners’ equipment.
High Quality
Components
From boards and modules to standalone devices, Nonin helps OEMs develop reliable products for use in sleep medicine, hospital patient monitoring, emergency medicine, telehealth, and outpatient monitoring applications.
Strategic
Partnerships
With high quality solutions and a non-infringing patent portfolio, Nonin is much more than a parts supplier. We consider ourselves a strategic partner to our OEM customers, and are always looking for ways to support their new product innovations. We offer our partners product development support as well as clinical and technical integration resources.
Exclusive Technologies
A unique approach to oximetry monitoring
Nonin Medical Port Devices Driver Download Win 7
Only Nonin Medical offers PureSAT®, PureLight®, and SenSmart® capabilities. These exclusive technologies provide considerable advantages for our OEM partners.
Multiple
Parameters
The SenSmart 3500SP accepts Nonin sensors for pulse oximetry (SpO2), cerebral and tissue oximetry (rSO2), carboxyhemoglobin (COHb), and methemoglobin (MetHb).
Note: Nonin sensors with COHb and MetHb capability are not yet cleared for EMS use in the United States.
PureSAT
Signal Processing
Nonin PureSAT technology uses intelligent pulse-by-pulse filtering to provide precise oximetry measurements—even in the presence of motion, low perfusion or other challenging conditions.
Nonin Medical Port Devices Driver Download
SenSmart
Regional Oximetry
SenSmart sensor and signal processing technologies feature a proprietary dual emitter/dual receiver sensor architecture and four wavelengths of light to enhance the accuracy of rSO2 measurements.
PureLight
Sensor Technology
All Nonin pulse oximetry sensors use our proven PureLight LEDs that emit a single, pure and clean light source; accurate from patient to patient and sensor to sensor.
Nonin Advantages
Low power—Our approach to signal processing reduces power consumption requirements in most applications.
Ease of integration—Nonin OEM solutions are designed for easy and flexible integration.
Reliability—Studies have shown that Nonin oximetry and sensors provide consistent results with a long service life, helping maximize the value of our OEM partners’ products.
Accuracy—Nonin products provide unmatched accuracy. Nonin is one of the only pulse oximetry manufacturers with clinical claims for both low perfusion and motion.
Support—We support our OEM partners from design to integration, testing, regulatory applications and launch. We also provide tools and training.
Components for OEMs
Wireless
Products
Nonin wireless products include the 3230 Bluetooth® LE finger pulse oximeter and the WristOx2® Model 3150 with BLE wrist-worn pulse oximeter.
Nonin Medical Port Devices Driver Download Windows 10
External
Modules
External pulse oximeter modules like the Nonin Xpod® connect to medical devices through an external port such as mini-12 or USB.
Internal
Modules
Nonin Medical Port Devices Driver Downloads
The Nonin OEM III Pulse Oximetry Module is integrated in sleep diagnostics systems, monitors, or ventilators.
Nonin OEM Development Kits
Evaluate our OEM solutions with a development kit that includes the product, demonstration software, technical documentation, and any additional interface components. Try one of our development kits to quickly and easily see how Nonin OEM solutions can meet your monitoring needs.
Contact Nonin’s OEM Sales Team
WristOx2® Model 3150 OEM with Bluetooth® Low Energy
OEM
Xpod® 3012LP
OEM
SenSmart® 3500SP
OEM
Model 3230 Bluetooth® Low Energy
Fingertip
Model 3231 USB
Fingertip